Z-2056-2021 Class II Ongoing

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-2056-2021 was initiated by GE Healthcare, LLC on Jun 4, 2021 and is designated Class II. Reason for recall: GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX"… The recall status is ongoing. Affected quantity: 248 units.

Recall details

Product Type
Devices
Report Date
Jul 21, 2021
Initiation Date
Jun 4, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
248 units

Product description

Discovery MR750 3.0T - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

Reason for recall

GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.

Distribution pattern

Worldwide distributions.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2056-2021
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for GE Healthcare, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for GE Healthcare, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about GE Healthcare, LLC →

Source: openFDA record for Z-2056-2021. Report a data issue: hello@regdatalab.com.