Z-2059-2013 Class II Terminated

Recalled by Philips Healthcare Inc. — Andover, MA

FDA device recall Z-2059-2013 was initiated by Philips Healthcare Inc. on Aug 8, 2013 and is designated Class II. Reason for recall: The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system. The recall status is terminated (terminated Sep 3, 2020). Affected quantity: 89.

Recall details

Product Type
Devices
Report Date
Sep 4, 2013
Initiation Date
Aug 8, 2013
Termination Date
Sep 3, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
89

Product description

Philips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.

Reason for recall

The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.

Distribution pattern

Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, GA, MO, MS, NC, NH, NY, PA and TX., and the countries of Canada, Australia, China, Czech Republic, France, Germany, Greece, Italy, Netherlands, Portugal, Russia, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates and United Kingdom.

Code information

437765/SN09000009, 452854/SN10000003, 463879/SN10000016, 500643/SN12000006,504710/SN12000011, 511454/SN12000018, 519869/SN13000005, 524432/SN13000007, 500643/SN12000006, 436570/SN09000008, 511492/SN12000019, SN09020039, SN09010077, 438946/SN09000011, SN10002064, SN11020126, SN12020282, 516048/SN12000021, SN13020163, 524411/SN13000011, 409205/SN08000002, 438474/SN09000040, 448425/SN10000001, 449743/SN10000010, 461559/SN10000014, 493489/SN11000008, 496231/SN12000003, 434381/SN09000005, 456026/SN10000011, 465845/SN11000005, 471608/SN11000001, 481912/SN11000004, 508150/SN12000015, 524367/SN13000009, 524510/SN13000010, SN07020246, 485916/SN11000007, 428148/SN09000006, 423585/SN09000003, 500351/SN12000009, 506215/SN12000012, 520201/SN13000003, 409442/SN08000003, 419279/SN09000001, 518054/SN13000001, 460532/SN10000013, SN09020304, 443304/SN09000012, 443310/SN09000013, 416734/SN08000005, 420531/SN09000002, 423195/SN09000004, 460544/SN10000017, 454997/SN10000009, 496323/SN12000002, 496324/SN12000001, 399965/SN08000001, 438232/SN09000010, SN09020302, 443840/SN09000014, 452406/SN10000004, 454310/SN10000005, 478444/SN11000003, 502350/SN12000007, 502355/SN12000016, 520759/SN13000006, 434429/SN09000007, 451666/SN10000002, 453781/SN10000008, 453960/SN10000007, 459266/SN10000012, 460755/SN10000015, 466174/SN10000019, 468076/SN10000020, 471702/SN11000002, 499486/SN12000004, 501950/SN12000008, 503132/SN12000010, 507693/SN12000013, 507694/SN12000014, 519994/SN13000004, 520759/SN13000006,SN09001033, SN09020312, SN09000826, SN11000761, 493873/SN11000009, SN12020117, SN12020322, SN12020440

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2059-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Healthcare Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Healthcare Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Healthcare Inc. →

Source: openFDA record for Z-2059-2013. Report a data issue: hello@regdatalab.com.