Z-2086-2013 Class II Terminated

Recalled by Alphatec Spine, Inc. — Carlsbad, CA

FDA device recall Z-2086-2013 was initiated by Alphatec Spine, Inc. on Jun 28, 2013 and is designated Class II. Reason for recall: The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a… The recall status is terminated (terminated Oct 9, 2014). Affected quantity: 333 screws.

Recall details

Product Type
Devices
Report Date
Sep 4, 2013
Initiation Date
Jun 28, 2013
Termination Date
Oct 9, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
333 screws

Product description

Illico Ti Cannulated Polyaxial Screw 7.5mm X 40mm. The Illico MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development.

Reason for recall

The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a potential risk that when the screw is mated to a polyaxial screwdriver, the guide wire may not pass through the screwdriver.

Distribution pattern

Worldwide distribution: US (nationwide) including states of:US states: AZ, CA, FL, GA, IL, MD, NJ, NY, OK, PA, TN, TX,and WA ; and countries of: Spain, France, Japan. International: Spain, France, Japan.

Code information

Part Number 73875-40. Lots Affected: 647289, 648923, 649244, 649556, 649813, 650040, 650262, 650589.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2086-2013
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Alphatec Spine, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2086-2013. Report a data issue: hello@regdatalab.com.