Z-2129-2016 Class II Terminated
FDA device recall Z-2129-2016 was initiated by Medtronic Navigation on Jun 20, 2016 and is designated Class II. Reason for recall: Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user. The recall status is terminated (terminated Jun 21, 2017). Affected quantity: 52.
Recall details
- Product Type
- Devices
- Report Date
- Jul 13, 2016
- Initiation Date
- Jun 20, 2016
- Termination Date
- Jun 21, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 52
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - US Nationwide.