Z-2131-2013 Class III Terminated

Recalled by Copan Italia — Brescia, N/A

FDA device recall Z-2131-2013 was initiated by Copan Italia on Aug 9, 2013 and is designated Class III. Reason for recall: The recall of the WASP (Walk Away Specimen Process) instrument was initiated because Copan Italia SpA has upgraded the software concerning the broth inoculation table feature. The upgrade has been… The recall status is terminated (terminated Dec 12, 2013). Affected quantity: 63 units.

Recall details

Product Type
Devices
Report Date
Sep 11, 2013
Initiation Date
Aug 9, 2013
Termination Date
Dec 12, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
63 units

Product description

WASP (Walk Away Specimen Process). Automatic Planting and Streaking Instrument Catalog number: W086 WASP is a fully automated system used to process microbiological specimens in liquid phase for the investigation of infectious microbes.

Reason for recall

The recall of the WASP (Walk Away Specimen Process) instrument was initiated because Copan Italia SpA has upgraded the software concerning the broth inoculation table feature. The upgrade has been considered necessary from the manufacturer in order to eliminate the very low risk of rotation of the broth table when the door is open.

Distribution pattern

Worldwide Distribution - USA including Oregon and California., and the countries of France, Belgium, Italy, Denmark, Norway, Switzerland, Romania, Finland, Germany, South Africa, Canada, Ireland, UK, Sweden and Japan

Code information

Serial Numbers: W086-61-084/106/115/119/134/136; W086-050-179; W086-61-087/109/125/127/130/133; W086-050-144/151/153/157/170/171/181/184; W086-61-088/097/100/117; W086-050-177/178/183; W086-61-093; W086-050-143/155; W086-61-096; W086-050-164/169/159/175; W086-61-103/137/141/142; W08661-104/114118/128/138; W086050-182; W08661 111; W08661 113; W08661 124; W08661129/140; W086050150/152/154/185; W08661 139; W086050 147; W08661 110; W086050160/161/162; W086050 180; W086050 187.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2131-2013
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Copan Italia? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2131-2013. Report a data issue: hello@regdatalab.com.