Z-2440-2021 Class II Terminated
Recalled by Copan Italia
— Brescia, N/A
FDA device recall Z-2440-2021 was initiated by Copan Italia on Jul 9, 2021 and is designated Class II. Reason for recall: A sterility assurance level of 10-6 cannot be guaranteed due to intentional fraud and data integrity issues at the contract sterilizer, Steril Milano. The recall status is terminated (terminated Jan 25, 2024). Affected quantity: 3,029,400 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 15, 2021
- Initiation Date
- Jul 9, 2021
- Termination Date
- Jan 25, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3,029,400 units
Product description
COPAN FLOQSwabs, STERILE.EO, for drawing clinical samples. CE0123, for the following reference numbers REF:501CS01, Minitip, sterile Single Wrapped, Molded bp 80mm; REF: 20226, P/N 501CS01Q100, Minitip, Sterile, Single Wrapped, bp 60MM, BUNDLES OF 20 pcs; REF: 501CS01.Q20, P/N 1127700, Minitip Applicator Sterile Ind. wrapped., bp 60mm, bundles of 20; REF: 502CS01, Regular, Sterile single Wrapped, Molded bp 80mm; REF: 503CS01, Flexible, Sterile Single Wrapped, Molded bp 100mm; REF: 220252, P/N 503CS01.BD, Flexible, Sterile Single Wrapped Molded bp 100 mm; REF: 503CS01/503CS01.BX, Flexible, Sterile Single Wrapped, Molded bp 100mm REF: 50U004DS, hDNA Free, Regular sterile in dry tube Molded Break Point 20 mm; REF: 50U009DS02, hDNA Free, Regular, Sterile, 2pcs Packed; REF: 518CS01, Minitip, Ster. Single Wrapped, Molded bp 100mm; REF: 220251, P/N: 518CS01.BD, Minitip, sterile Single Wrapped Molded bp 100mm; REF: 519C/519C.BX, Regular Sterile, Molded break point 100 mm; REF: 220250 P/N: 519CS01.BD, Regular, Sterile Single wrapped Molded bp 100mm; REF: 525CS01, Ultra-Thin Minitip, sterile single wrapped, molded bp 80 mm; REF: 528C, Large Tip Swab, Handle with Stopper, in dry tube REF: 534CS01, Flexible Nylon, Sterile Single Wrapped; REF: 551C, Minitip Sterile in dry tube, Molded Break Point 80 mm; REF: 552C/552C.BX, Regular sterile in dry tube Molded Break Point 80 mm; REF: 553C, Flexible sterile in dry tube, molded bp 100 mm; REF: 553C.LC, Flexible Sterile in dry tube, molded bp 100 mm; REF: 5U002S, Regular, sterile, in tube, 80 mm breakpoint, with low stopper; 5U048N01.BIO, REF: 5U055S02.HG, Regular FLOQSwabs with 20mm breakpoint 100 peelpouches, 2 swabs per peelpouch; REF: 5U056S.IUO, FLOQSwats ADULT IN DRY TUBE, Investigational Use Only; REF: 5UO57S.IUO, FLOQSwabs CHILD IN DRY TUBE, Investigational Use Only
Reason for recall
A sterility assurance level of 10-6 cannot be guaranteed due to intentional fraud and data integrity issues at the contract sterilizer, Steril Milano.
Distribution pattern
US nationwide distribution, including Puerto Rico.
Code information
Mode Number: Lot number: 501CS01: 1816581 and 2027890; 501CS01.Q100: PJ10453, PJ10636 and PJ10667; 501CS01.Q20: J10407; 502CS01: 1816964, 2027239, 2027877; 503CS01: 1807589, 1809095, 1814761, 1814762, 1814763, 1816968, 1821362, 2025887; 503CS01.BD: 2031801; 503CS01.BX: 50U004DS and 1818685; 50U009DS02: 1825274; 518CS01: 2008058 and 2027915; 518CS01.BD: 2008058; 519C: 1816921; 519C.BX: 1811060 and 1816921; 519CS01.BD: 2027871; 525CS01: 2009380; 528C: 1808601, 534CS01, 1707256; 551C: 1813221; 552C: 1819144; 552C.BX: 1819144; 553C: 1810293, 1816440, 1817638, 2004854: 553C.LC: 1811976, 1814054, 1814717, 1817637, 1820999; 5U002S: 1812705; 5U048N01.BIO: 1810727 and 1818912; 5U055S02.HG: 1814413; 5U056S.IUO: 1816448; 5U057S.IUO: 1816451
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2440-2021
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Copan Italia? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works