Z-2139-2020 Class II Terminated
FDA device recall Z-2139-2020 was initiated by Radiometer Medical ApS on Nov 21, 2019 and is designated Class II. Reason for recall: The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the… The recall status is terminated (terminated Nov 8, 2024). Affected quantity: 1233 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 3, 2020
- Initiation Date
- Nov 21, 2019
- Termination Date
- Nov 8, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1233 units
Product description
Reason for recall
Distribution pattern
Worldwide distribution - US Nationwide. Also distributed OUS.