Z-2139-2020 Class II Terminated

Recalled by Radiometer Medical ApS — Bronshoj, N/A

Recall Details

Product Type
Devices
Report Date
June 3, 2020
Initiation Date
November 21, 2019
Termination Date
November 8, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1233 units

Product Description

ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Reason for Recall

The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.

Distribution Pattern

Worldwide distribution - US Nationwide. Also distributed OUS.

Code Information

ABL90 FLEX Model no. 393-090: R0659 - R0977 and Spare Part Hemolyzer Units (902-980): R0526 - R0941