Z-2142-2014 Class III Terminated

Recalled by Carestream Health Inc. — Rochester, NY

Recall Details

Product Type
Devices
Report Date
August 13, 2014
Initiation Date
July 2, 2014
Termination Date
May 25, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1633 boxes

Product Description

INSIGHT Dental Film, Catalog Number/REF 811 0785, KODAK and Carestream DENTAL labels --- Carestream Health Inc.

Reason for Recall

The Indicator Dot on the film packet's outer printed paper labeling was incorrectly positioned relative to the location of the raised dot on the dental intraoral film. This may cause confusion for the dental practitioner. If the processed film is placed on the view box by referencing the film's raised dot, the image orientation will be rotated 180 degrees.

Distribution Pattern

Worldwide Distribution.

Code Information

Lot Numbers: 53901201, 53801303, 53801303-1, 54601201, 54601201-1