Z-2143-2016 Class III Terminated
FDA device recall Z-2143-2016 was initiated by Ortho-Clinical Diagnostics on May 16, 2016 and is designated Class III. Reason for recall: If the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numerical results fall between Negative and Reactive, the IM incorrectly displays the interpretive text as… The recall status is terminated (terminated Jun 8, 2018). Affected quantity: US: 30 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 20, 2016
- Initiation Date
- May 16, 2016
- Termination Date
- Jun 8, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- US: 30 units
Product description
enGen Laboratory Automation System, Product Code engen, Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; IVD
Reason for recall
If the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numerical results fall between Negative and Reactive, the IM incorrectly displays the interpretive text as Borderline and sends the incorrect text to the LIS. The IFU states the interpretation term for these numerical result values is Retest?.¿ The difference in terminology exists for US Markets only.
Distribution pattern
Nationwide
Code information
Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; Serial Numbers/J Numbers: JIM341866, JIM341820, JIM341679, JIM341352, JIM343421, JIM341174, JIM336745, JIM340667, JIM338283, JIM-341592, JIM336957, JIM341996, JIM342334, JIM343694, JIM336855, JIM334236, JIM341415, JIM341392, JIM336481, JIM340934, JIM336167, JIM342249, JIM338290, JIM341750, JIM341426, JIM334986, JIM334527, JIM340024, JIM341021, JIM340658