Z-2145-2016 Class III Terminated
FDA device recall Z-2145-2016 was initiated by ORTHO-CLINICAL DIAGNOSTICS on May 9, 2016 and is designated Class III. Reason for recall: Ortho Clinical Diagnostics (OCD) confirmed that specific lots of VITROS Immunodiagnostic Products AFP Reagent Pack product exhibit an increase in the reaction signal over time potentially resulting in… The recall status is terminated (terminated Jun 8, 2018). Affected quantity: US: Lot 6471 - 282, Lot 6480 - 344.
Recall details
- Product Type
- Devices
- Report Date
- Jul 20, 2016
- Initiation Date
- May 9, 2016
- Termination Date
- Jun 8, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- US: Lot 6471 - 282, Lot 6480 - 344
Product description
VITROS Immunodiagnostic Products AFP Reagent Pack, REF/Product Code 680 0784, Unique Device Identifier No. 10758750000876, IVD; --- Ortho-Clinical Diagnostics Percoed, Bridgend, UK --- Please note: the VITROS Immunodiagnostic Products AFP Calibrators, Unique Device Identifier No. 10758750000913, REF 680 0863, were included in the communication because they are lot-linked to the Reagent Packs and cannot be used with any other lot of reagent. There is no known failure with the Calibrators. For the quantitative measurement of alpha-fetoprotein (AFP) concentrations in human serum using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System to aid in the management of patients with non-seminomatous testicular cancer.
Reason for recall
Ortho Clinical Diagnostics (OCD) confirmed that specific lots of VITROS Immunodiagnostic Products AFP Reagent Pack product exhibit an increase in the reaction signal over time potentially resulting in the inability to calibrate and use the product.
Distribution pattern
US Nationwide Distribution
Code information
Lots 6471 (expiry date 19-Jul-2016), 6480 (expiry date 23-Aug-2016)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2145-2016
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab