Z-2164-2026 Class II Ongoing
FDA device recall Z-2164-2026 was initiated by Dexcom, Inc. on Apr 14, 2026 and is designated Class II. Reason for recall: A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the… The recall status is ongoing. Affected quantity: 58582.
Recall details
- Product Type
- Devices
- Report Date
- May 20, 2026
- Initiation Date
- Apr 14, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 58582
Product description
Reason for recall
Distribution pattern
SW12300 & SW13355 Dexcom G7 iOS and watchOS app versions 2.9.0 through 2.11.2: Worldwide - US Nationwide including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Austria, Australia, Belgium, Bahrain, Canada, Czech Republic, Germany, Denmark, Estonia, Finland, France, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, Kuwait, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom. SW14245 Dexcom ONE iOS App versions 1.5.0 and 1.6.0 - Only International distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Czech Republic, Estonia, France, Greece, Croatia, Ireland, Italy, Lithuania, Latvia, Malaysia, Monaco, Netherlands, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Spain, Sweden, United Kingdom.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2164-2026
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026