Z-2448-2025 Class I Ongoing

Recalled by Dexcom, Inc. — San Diego, CA

FDA device recall Z-2448-2025 was initiated by Dexcom, Inc. on Jul 24, 2025 and is designated Class I. Reason for recall: The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS… The recall status is ongoing. Affected quantity: 907842.

Recall details

Product Type
Devices
Report Date
Sep 10, 2025
Initiation Date
Jul 24, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
907842

Product description

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used of each are the same for each user. The watchOS app cannot be installed independently from the iOS phone app Component: Dexcom G7 CGM iOS App

Reason for recall

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

Distribution pattern

SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the countries of Andorra, United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW12300 G7 iOS CGM App Worldwide Distribution - US Nationwide and the countries of United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW14245 Dexcom ONE+ Android Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Israel, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden SW14244 D1G7 iOS Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden

Code information

Lot Code: UDI - SW1355- 00386270005168

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2448-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Dexcom, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Dexcom, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Dexcom, Inc. →

Source: openFDA record for Z-2448-2025. Report a data issue: hello@regdatalab.com.