Z-2206-2016 Class II Terminated

Recalled by RAYSEARCH LABORATORIES AB — Stockholm, N/A

FDA device recall Z-2206-2016 was initiated by RAYSEARCH LABORATORIES AB on Jul 1, 2016 and is designated Class II. Reason for recall: An issue was found with the evaluation of biological clinical goals in RayStation 3.0, RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7 and RayStation 5. Biological clinical goals for an… The recall status is terminated (terminated Sep 25, 2017). Affected quantity: US: 170 units.

Recall details

Product Type
Devices
Report Date
Jul 27, 2016
Initiation Date
Jul 1, 2016
Termination Date
Sep 25, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US: 170 units

Product description

RayStation 3.0, RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7 and RayStation 5 --- Radiation Therapy Treatment Planning System --- designed for treatment planning and analysis of radiation therapy, provides treatment unit set up parameters and estimates dose distributions.

Reason for recall

An issue was found with the evaluation of biological clinical goals in RayStation 3.0, RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7 and RayStation 5. Biological clinical goals for an adapted plan based on another planning CT than the original planning CT will show incorrect values. -- To the best of the firm's knowledge, the issue has not caused any patient mistreatment or other incidents. However, the user must be aware of the following information to avoid incorrect plan evaluation during treatment planning.

Distribution pattern

California, Connecticut, Delaware, Florida, Hawaii, Maine, Missouri, New York, Ohio, Texas and Washington

Code information

Software build numbers: 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.42, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.4.4, 4.7.5.4, 5.0.0.37, 5.0.1.11 and 5.0.2.35

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2206-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for RAYSEARCH LABORATORIES AB? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for RAYSEARCH LABORATORIES AB: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about RAYSEARCH LABORATORIES AB →

Source: openFDA record for Z-2206-2016. Report a data issue: hello@regdatalab.com.