Z-2208-2012 Class II Terminated

Recalled by Philips Healthcare Inc. — Andover, MA

FDA device recall Z-2208-2012 was initiated by Philips Healthcare Inc. on Jul 11, 2012 and is designated Class II. Reason for recall: Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors. The recall status is terminated (terminated Jan 23, 2017). Affected quantity: 12 units.

Recall details

Product Type
Devices
Report Date
Aug 22, 2012
Initiation Date
Jul 11, 2012
Termination Date
Jan 23, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product description

Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.

Reason for recall

Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.

Distribution pattern

Worldwide Distribution - USA including IN, MA, MI, MO, NC, NY PA, SC, and WI and to the countries of Canada, HongKong, New Zealand.

Code information

Serial Numbers: SN07000141 SN07000109 SN5000162 SN5000421 SN07000667 200148 200115 200081 300114

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2208-2012
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Healthcare Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Healthcare Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Healthcare Inc. →

Source: openFDA record for Z-2208-2012. Report a data issue: hello@regdatalab.com.