Z-2208-2012 Class II Terminated
FDA device recall Z-2208-2012 was initiated by Philips Healthcare Inc. on Jul 11, 2012 and is designated Class II. Reason for recall: Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors. The recall status is terminated (terminated Jan 23, 2017). Affected quantity: 12 units.
Recall details
- Product Type
- Devices
- Report Date
- Aug 22, 2012
- Initiation Date
- Jul 11, 2012
- Termination Date
- Jan 23, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 12 units
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - USA including IN, MA, MI, MO, NC, NY PA, SC, and WI and to the countries of Canada, HongKong, New Zealand.