Z-2224-2012 Class II Terminated
FDA device recall Z-2224-2012 was initiated by Biomet U.K., Ltd. on Jul 27, 2012 and is designated Class II. Reason for recall: A drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model remained the same thickness but the problem dimensions were accidentally struck… The recall status is terminated (terminated Feb 15, 2013). Affected quantity: 119 units.
Recall details
- Product Type
- Devices
- Report Date
- Aug 29, 2012
- Initiation Date
- Jul 27, 2012
- Termination Date
- Feb 15, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 119 units
Product description
Reason for recall
Distribution pattern
Worldwide Distribution-USA-including the states of IN, SD, and TX and the countries of Austria, China, Japan, Netherlands, Sweden, Turkey, and UK.