Z-2248-2017 Class II Terminated

Recalled by Stryker Communications — Flower Mound, TX

FDA device recall Z-2248-2017 was initiated by Stryker Communications on Apr 21, 2017 and is designated Class II. Reason for recall: The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount… The recall status is terminated (terminated Sep 20, 2017). Affected quantity: 57 units.

Recall details

Product Type
Devices
Report Date
Jun 14, 2017
Initiation Date
Apr 21, 2017
Termination Date
Sep 20, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
57 units

Product description

Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated

Reason for recall

The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.

Distribution pattern

Worldwide Distribution - US and Canada

Code information

Serial Numbers: 7735170-X68377, 7735170-X68397, 7735170-X14846, 7735170-X15688, 7735170-X14492, 7735170-X14495, 7735170-X14497, 7735170-X14501, 7735170-X14502, 7735170-X14504, 7735170-X14506, 7735170-X14507, 7735170-X14508, 7735170-X14510, 7735170-X68383, 7735170-X68384, 7735170-X68385, 7735170-X68386, 7735170-X68393, 7735170-X68394, 7735170-X68395, 7735170-X68396, 7735170-X68392, 7735170-X68400, 7735170-X68401, 7735170-X15607, 7735170-X15608, 7735170-X15609, 7735170-X15610, 7735170-X15611, 7735170-X15612, 7735170-X15613, 7735170-X15614, 7735170-X15615

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2248-2017
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Stryker Communications? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2248-2017. Report a data issue: hello@regdatalab.com.