Z-2249-2017 Class II Terminated
Recalled by Alphatec Spine, Inc.
— Carlsbad, CA
FDA device recall Z-2249-2017 was initiated by Alphatec Spine, Inc. on May 5, 2017 and is designated Class II. Reason for recall: Alphatec Spine is recalling the Arsenal Spinal Fixation System Set Screw due to a trend in set screw postoperative disengagement from the screw body. The recall status is terminated (terminated Jun 6, 2017). Affected quantity: 16,959 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 14, 2017
- Initiation Date
- May 5, 2017
- Termination Date
- Jun 6, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 16,959 units
Product description
Arsenal Spinal Fixation System, Set Screw, Part No. 47027 Product Usage: The Arsenal Spinal Fixation System is intended for posterior, non-cervical, spinal fixation as an adjunct to fusion for the treatment of degenerative disease, deformity, and trauma indications. The Arsenal System consists of a variety of shapes and sizes of rods, screws, hooks, connectors, and bridges that provide temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. The screws, hooks, connectors, and bridges are manufactured from surgical grade titanium alloy (Ti-6AI-4V ELl). The rods are available in commercially pure titanium, titanium alloy, and cobalt chrome (CP Ti Grade 4, Ti-6AI-4V ELl, and Co-28Cr-6Mo).
Reason for recall
Alphatec Spine is recalling the Arsenal Spinal Fixation System Set Screw due to a trend in set screw postoperative disengagement from the screw body.
Distribution pattern
Worldwide Distribution - US Nationwide in the states of (TX, AL, UT, FL, IA, SD, NV, OH, CA, MN, CO, AZ, OK, MI, VA, KY, and Puerto Rico) and country of Japan
Code information
671949 681640 684646 687843 690206 692720 693064 671948 681641 684646 687843 690206 692705 674136 681642 684647 687847 690210 692718 674137 682061 686374 687847 691848S1 692719 674431 682060 684645 689847 691848 692855 674432 682062 686372 689847 691849 692855 674433 682063 686375 689849 692597 692706 671950 684429 687839 690127 692597 692864 674429 684430 687839 689848 692364 692910 674430 684432 687840 690207 692598 692911 681638S1 684644 687840 690207 691590 692911 681638 684433 687842 690208 692599 693065 681639 684431 687845 690209 692704 693057
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2249-2017
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.