Z-2251-2020 Class II Terminated

Recalled by Abbott Laboratories — Irving, TX

FDA device recall Z-2251-2020 was initiated by Abbott Laboratories on Apr 2, 2020 and is designated Class II. Reason for recall: In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could… The recall status is terminated (terminated Apr 28, 2021). Affected quantity: 11.

Recall details

Product Type
Devices
Report Date
Jun 10, 2020
Initiation Date
Apr 2, 2020
Termination Date
Apr 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11

Product description

Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate sample handling and processing in the clinical laboratory. The system allows consolidations of multiple clinical chemistry and immunoassay analytical instruments into a unified workstation.

Reason for recall

In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results.

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of NC, ND, SC, TN. The countries of United Kingdom, France, New Zealand, Pakistan, Turkey.

Code information

All serial numbers in the field

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2251-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Laboratories? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Abbott Laboratories: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Abbott Laboratories →

Source: openFDA record for Z-2251-2020. Report a data issue: hello@regdatalab.com.