Z-2265-2014 Class II Terminated

Recalled by Synthes, Inc. — West Chester, PA

FDA device recall Z-2265-2014 was initiated by Synthes, Inc. on Jun 30, 2014 and is designated Class II. Reason for recall: A screw set was returned from the field for destruction after being exposed to flood water, which was erroneously reprocessed and returned to the field. Initial investigative testing indicated the… The recall status is terminated (terminated Dec 18, 2015). Affected quantity: 4.

Recall details

Product Type
Devices
Report Date
Aug 27, 2014
Initiation Date
Jun 30, 2014
Termination Date
Dec 18, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4

Product description

Synthes 3.0 MM TI Cannulated Screw Instrument and Implant Set

Reason for recall

A screw set was returned from the field for destruction after being exposed to flood water, which was erroneously reprocessed and returned to the field. Initial investigative testing indicated the components within the set may be cytotoxic.

Distribution pattern

Distributed in the states of WI and MI.

Code information

Part Number: 145.17E Lot Number: EV01609, individual part numbers contained within 145.17E: 304.49 309.501 310.221 310.804 311.43 312.151 312.153 313.969 314.463 314.464 314.465 319.291 319.292 319.702 319.97 398.408 398.409 402.608-402.640 402.714-402.740 419.89 419.972 292.622, and 292.623.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2265-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.