Z-2276-2013 Class II Terminated
FDA device recall Z-2276-2013 was initiated by Elekta, Inc. on Jul 2, 2013 and is designated Class II. Reason for recall: Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values. The recall status is terminated (terminated May 8, 2014). Affected quantity: 45.
Recall details
- Product Type
- Devices
- Report Date
- Oct 2, 2013
- Initiation Date
- Jul 2, 2013
- Termination Date
- May 8, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 45
Product description
Monaco RTP System Radiation treatment planning
Reason for recall
Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values.
Distribution pattern
Nationwide Distribution including AK, CA, CO, CT, FL, IL, IN, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, SC, TN, TX, WA and Puerto Rico.
Code information
Versions 3.10.00 and later
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2276-2013
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab