Z-2283-2019 Class II Ongoing

Recalled by Luminex Corporation — Austin, TX

FDA device recall Z-2283-2019 was initiated by Luminex Corporation on Feb 20, 2019 and is designated Class II. Reason for recall: Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Aug 21, 2019
Initiation Date
Feb 20, 2019
Voluntary/Mandated
Voluntary: Firm initiated

Product description

ARIES HSV 1&2 Assay , REF 50-10017, UDI # 00840487100295

Reason for recall

Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.

Distribution pattern

US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom

Code information

Lot numbers & discontinue use by dates for Part Number 50-10017: AA7308 - Immediately; AA7485 - 2/23/2019; AA7545 - 3/13/2019; AA7625 - 4/10/2019; AA7630 - 3/16/2019; AA7636 - 4/3/2019; AA7945 - 4/25/2019; AA7985 - 5/8/2019

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2283-2019
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.