Z-2289-2020 Class II Terminated

Recalled by Elekta, Inc. — Atlanta, GA

FDA device recall Z-2289-2020 was initiated by Elekta, Inc. on Apr 30, 2020 and is designated Class II. Reason for recall: potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray. The recall status is terminated (terminated Aug 4, 2021). Affected quantity: 32 units.

Recall details

Product Type
Devices
Report Date
Jun 17, 2020
Initiation Date
Apr 30, 2020
Termination Date
Aug 4, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
32 units

Product description

Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system - Product Usage: is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

Reason for recall

potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray.

Distribution pattern

Worldwide distribution - US Nationwide distribution including in the states of PA, WI, NY, TN, NJ, IA, TX, and the countries of Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Thailand, Turkey, UK.

Code information

Serial numbers 600001 - 600043

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2289-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2289-2020. Report a data issue: hello@regdatalab.com.