Z-2289-2020 Class II Terminated
FDA device recall Z-2289-2020 was initiated by Elekta, Inc. on Apr 30, 2020 and is designated Class II. Reason for recall: potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray. The recall status is terminated (terminated Aug 4, 2021). Affected quantity: 32 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 17, 2020
- Initiation Date
- Apr 30, 2020
- Termination Date
- Aug 4, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 32 units
Product description
Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system - Product Usage: is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.
Reason for recall
potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray.
Distribution pattern
Worldwide distribution - US Nationwide distribution including in the states of PA, WI, NY, TN, NJ, IA, TX, and the countries of Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Thailand, Turkey, UK.
Code information
Serial numbers 600001 - 600043