Z-2303-2013 Class II Terminated

Recalled by Radiometer America Inc — Westlake, OH

FDA device recall Z-2303-2013 was initiated by Radiometer America Inc on Oct 8, 2012 and is designated Class II. Reason for recall: RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the… The recall status is terminated (terminated Dec 24, 2013). Affected quantity: 13,632 boxes/4 membranes.

Recall details

Product Type
Devices
Report Date
Oct 9, 2013
Initiation Date
Oct 8, 2012
Termination Date
Dec 24, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,632 boxes/4 membranes

Product description

Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.

Reason for recall

RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.

Distribution pattern

Nationwide Distribution, including the states of: AK, AL, AR, AZ, BM, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WV.

Code information

Part Number: 942-066 R1741 to R1869

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2303-2013
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Radiometer America Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Radiometer America Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Radiometer America Inc →

Source: openFDA record for Z-2303-2013. Report a data issue: hello@regdatalab.com.