Z-2308-2025 Class II Ongoing

Recalled by GE Medical Systems, LLC — Waukesha, WI

FDA device recall Z-2308-2025 was initiated by GE Medical Systems, LLC on Jun 20, 2025 and is designated Class II. Reason for recall: GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T… The recall status is ongoing. Affected quantity: 56 units.

Recall details

Product Type
Devices
Report Date
Aug 20, 2025
Initiation Date
Jun 20, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
56 units

Product description

GE SIGNA Hero, Nuclear Magnetic Resonance Imaging System

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Distribution pattern

Worldwide distribution.

Code information

GTIN: 00195278472083, 00195278486813, 001952784868131, Not Applicable; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2308-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2308-2025. Report a data issue: hello@regdatalab.com.