Z-2324-2015 Class II Terminated
FDA device recall Z-2324-2015 was initiated by Medtronic Navigation, Inc. on Jul 17, 2015 and is designated Class II. Reason for recall: The adhesive between the sphere assembly and the biopsy needle shaft may be compromised, causing the sphere assembly to move on the needle shaft. This may result in inaccurate navigation and improper… The recall status is terminated (terminated Apr 7, 2016). Affected quantity: 1415 total.
Recall details
- Product Type
- Devices
- Report Date
- Aug 12, 2015
- Initiation Date
- Jul 17, 2015
- Termination Date
- Apr 7, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1415 total
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - US (nationwide) and Canada.