Z-2325-2015 Class II Terminated

Recalled by Medtronic Navigation, Inc. — Louisville, CO

FDA device recall Z-2325-2015 was initiated by Medtronic Navigation, Inc. on Jul 17, 2015 and is designated Class II. Reason for recall: The adhesive between the sphere assembly and the biopsy needle shaft may be compromised, causing the sphere assembly to move on the needle shaft. This may result in inaccurate navigation and improper… The recall status is terminated (terminated Apr 7, 2016). Affected quantity: 1415 total.

Recall details

Product Type
Devices
Report Date
Aug 12, 2015
Initiation Date
Jul 17, 2015
Termination Date
Apr 7, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1415 total

Product description

Medtronic Passive Biopsy Needle Kit, Part Number 9731754 - Passive Biopsy Fee Per Use (FPU) Kit. The image guided Passive Biopsy Needle is an accessory instrument with the StealthStation System. The Biopsy Needle is intended to be used for stereotaxy biopsy of brain tissue.

Reason for recall

The adhesive between the sphere assembly and the biopsy needle shaft may be compromised, causing the sphere assembly to move on the needle shaft. This may result in inaccurate navigation and improper placement of the needles biopsy window in the brain.

Distribution pattern

Worldwide Distribution - US (nationwide) and Canada.

Code information

LOT No: 0007525461. 0007599312

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2325-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Navigation, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Navigation, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Navigation, Inc. →

Source: openFDA record for Z-2325-2015. Report a data issue: hello@regdatalab.com.