Z-2343-2023 Class II Ongoing

Recalled by Olympus Corporation of the Americas — Center Valley, PA

Recall Details

Product Type
Devices
Report Date
August 16, 2023
Initiation Date
May 26, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
241 units

Product Description

Olympus Endoscope Reprocessor OER-Mini-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.

Reason for Recall

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

Distribution Pattern

Nationwide

Code Information

UDI-DI:- 04953170331619 Serial Numbers: 2300314 and the last three digits less than 314