Z-2349-2020 Class II Terminated

Recalled by RAYSEARCH LABORATORIES AB — Stockholm, N/A

Recall Details

Product Type
Devices
Report Date
June 17, 2020
Initiation Date
May 18, 2020
Termination Date
August 6, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

RayStation 9A, UDI # 07350002010174 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.

Reason for Recall

Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.

Distribution Pattern

US Nationwide distribution including in the states of Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.

Code Information

UDI # 07350002010174 Build number: 9.0.0.113