Z-2355-2018 Class II Terminated

Recalled by FHC, Inc. — Bowdoin, ME

Recall Details

Product Type
Devices
Report Date
July 11, 2018
Initiation Date
May 10, 2018
Termination Date
April 30, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2952 total

Product Description

microTargeting" Drive System Distributed by Medtronic, Product Number FC1006

Reason for Recall

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

Distribution Pattern

The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.

Code Information

UDI: 00873263001479