Z-0312-2025 Class I Ongoing

Recalled by FHC, Inc. — Bowdoin, ME

FDA device recall Z-0312-2025 was initiated by FHC, Inc. on Oct 4, 2024 and is designated Class I. Reason for recall: A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was… The recall status is ongoing. Affected quantity: 8 packs of 5ea (40 total).

Recall details

Product Type
Devices
Report Date
Nov 20, 2024
Initiation Date
Oct 4, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 packs of 5ea (40 total)

Product description

microTargeting Insertion Tube Set, P/N 66-IT-05P, Lot 244517. The intended use is to accurately guide a microelectrode or instrument in the brain during stereotactic functional neurosurgical procedures.

Reason for recall

A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.

Distribution pattern

Domestic distribution only to the following states: AZ, CA, PA, TN, TX, WA.

Code information

Package DI: 10873263004323; Contains DI Package: 00873263004326 LOT No. 244517

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0312-2025
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for FHC, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0312-2025. Report a data issue: hello@regdatalab.com.