Z-2359-2018 Class II Terminated

Recalled by FHC, Inc. — Bowdoin, ME

FDA device recall Z-2359-2018 was initiated by FHC, Inc. on May 10, 2018 and is designated Class II. Reason for recall: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one… The recall status is terminated (terminated Apr 30, 2019). Affected quantity: 2952 total.

Recall details

Product Type
Devices
Report Date
Jul 11, 2018
Initiation Date
May 10, 2018
Termination Date
Apr 30, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2952 total

Product description

microtargeting" Depth Stop Adapter, Product Model Number, Product Number 66-AC-DS(1.8),

Reason for recall

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

Distribution pattern

The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.

Code information

UDI: 00873263006375, 00873263001493

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2359-2018
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for FHC, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2359-2018. Report a data issue: hello@regdatalab.com.