Z-2356-2018 Class II Terminated
FDA device recall Z-2356-2018 was initiated by FHC, Inc. on May 10, 2018 and is designated Class II. Reason for recall: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one… The recall status is terminated (terminated Apr 30, 2019). Affected quantity: 2952 total.
Recall details
- Product Type
- Devices
- Report Date
- Jul 11, 2018
- Initiation Date
- May 10, 2018
- Termination Date
- Apr 30, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2952 total
Product description
Reason for recall
Distribution pattern
The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2356-2018
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for FHC, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.