Z-2357-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 11, 2018
- Initiation Date
- May 10, 2018
- Termination Date
- April 30, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2952 total
Product Description
microTargeting" STar" Drive System (Manual) Distributed by Medtronic, Product Number FC8001
Reason for Recall
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
Distribution Pattern
The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.
Code Information
UDI: 00873263001486, 00873263006351