Z-2357-2023 Class II Ongoing

Recalled by ORTHO-CLINICAL DIAGNOSTICS — Bridgend, N/A

Recall Details

Product Type
Devices
Report Date
August 16, 2023
Initiation Date
June 5, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,013 units

Product Description

VITROS Immunodiagnostic Products Troponin I ES Reagent Pack

Reason for Recall

Reagent packs could potentially contain incorrect wells, which will generate incorrect, and potentially false negative, results.

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Canada.

Code Information

Product Code: 6802301; UDI-DI: 10758750002504; Lot No.: 4800; Expiration Date: 10-Jun-2023.