Z-2363-2016 Class II Terminated
FDA device recall Z-2363-2016 was initiated by Merge Healthcare, Inc. on Feb 5, 2016 and is designated Class II. Reason for recall: When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same. The recall status is terminated (terminated Oct 25, 2016). Affected quantity: 49 sites potentially have the affected versions.
Recall details
- Product Type
- Devices
- Report Date
- Aug 17, 2016
- Initiation Date
- Feb 5, 2016
- Termination Date
- Oct 25, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 49 sites potentially have the affected versions
Product description
Merge PACS software. The firm name on the label is Merge Healthcare, Inc., Hartland, WI. Intended to create and display two-dimensional and three-dimensional images of anatomy from a series of digitally acquired images.
Reason for recall
When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same.
Distribution pattern
Worldwide Distribution -- USA and Australia.
Code information
Software versions 6.0.2.0 MR2, 6.0.3.0 MR3, 6.0.3.1 MR3 CU1