Z-2384-2012 Class II Terminated

Recalled by Philips Healthcare Inc. — Andover, MA

FDA device recall Z-2384-2012 was initiated by Philips Healthcare Inc. on Aug 14, 2012 and is designated Class II. Reason for recall: Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm. The recall status is terminated (terminated Jan 17, 2017). Affected quantity: 4 systems.

Recall details

Product Type
Devices
Report Date
Sep 26, 2012
Initiation Date
Aug 14, 2012
Termination Date
Jan 17, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 systems

Product description

Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications

Reason for recall

Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.

Distribution pattern

Worldwide Distribution - USA including OH and WA and the country of Canada

Code information

Lot or Serial Numbers With affected M-cabinet CXA; Serial number; 11-K0003; 12-B0001, 12-B0002; 12-B0003

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2384-2012
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Healthcare Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Healthcare Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Healthcare Inc. →

Source: openFDA record for Z-2384-2012. Report a data issue: hello@regdatalab.com.