Z-2398-2012 Class II Terminated

Recalled by Bio-Rad Laboratories — Redmond, WA

FDA device recall Z-2398-2012 was initiated by Bio-Rad Laboratories on Jul 27, 2012 and is designated Class II. Reason for recall: Prolonged exposure of the MONOLISA Anti-HBs EIA Conjugate Diluent component to light may result in elevated Optical Densities (OD) in some samples such that a negative sample could elevate to a grey… The recall status is terminated (terminated Dec 14, 2012). Affected quantity: 7,400 (approximately)- All lots since distribution date May 9, 2007.

Recall details

Product Type
Devices
Report Date
Sep 26, 2012
Initiation Date
Jul 27, 2012
Termination Date
Dec 14, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7,400 (approximately)- All lots since distribution date May 9, 2007

Product description

The Bio-Rad MONOLISA Anti-HBs EIA is for In Vitro Diagnostic Use and it has the catalog number 25220. It is a qualitative and quantitative enzyme immunoassay for the detection of antibody for hepatitis B surface antigen in human serum and EDTA or citrated plasma. The Bio-Rad MONOLISA Anti-HBs EIA is a qualitative and quantitative enzyme immunoassay for the detection of antibody for hepatitis B surface antigen in human serum and EDTA or citrated plasma. The assay results may be used as an aid in the determination of susceptibility to hepatitis B virus (HBV) infection in individuals prior to or following HBV vaccination or where vaccination status is unknown. Assay results may be used with other HBV serological markers for the laboratory diagnosis of HBV disease associated with HBV infection.

Reason for recall

Prolonged exposure of the MONOLISA Anti-HBs EIA Conjugate Diluent component to light may result in elevated Optical Densities (OD) in some samples such that a negative sample could elevate to a grey zone or low positive result.

Distribution pattern

Worldwide distribution: USA (nationwide) and countries of: Canada, South Africa, and Zimbabwe.

Code information

All lots distributed from May 5, 2007

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2398-2012
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Bio-Rad Laboratories? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Bio-Rad Laboratories: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Bio-Rad Laboratories →

Source: openFDA record for Z-2398-2012. Report a data issue: hello@regdatalab.com.