Z-2403-2020 Class II Terminated

Recalled by Merge Healthcare, Inc. — Hartland, WI

FDA device recall Z-2403-2020 was initiated by Merge Healthcare, Inc. on May 8, 2020 and is designated Class II. Reason for recall: An error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abnormality resulting… The recall status is terminated (terminated Jun 23, 2023). Affected quantity: 62 units.

Recall details

Product Type
Devices
Report Date
Jul 1, 2020
Initiation Date
May 8, 2020
Termination Date
Jun 23, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
62 units

Product description

Merge Healthcare Merge Application Server Software Release and Merge Healthcare Merge Cardio Workstation Software Release - Product Usage: VERICIS is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.

Reason for recall

An error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abnormality resulting in a false negative.

Distribution pattern

US Nationwide distribution including in the states of IL, TX, OR, AZ, OH, TN, CA, UT, MO, GA, NM, OK, VA, IN, CT, and VT.

Code information

Merge Cardio versions 10.X and 11.X with Knowledge Base versions 3.3, 3.4, 3.5 or 5.0 installed on or before February 21, 2020. Version 10.0 Pre-UDI Application server PN 88-00528-00; Workstation PN:88-00534-00 Version 10.1.2 Pre-UDI Application server PN 88-01075-00; Workstation PN 88-01078-00 Version 10.2 P2 Pre-UDI Application Server PN 88-01203-00 Version 10.2 P1&P2 Pre-UDI Software Release PN 87-00247-00 Version 10.3 UDI: (01) 00842000100041(10)10.3.0.1137(11)170104 Application server PN: 88-01233-00; Workstation PN 88-01239-00 Version 10.3 Patch 1 UDI: (01)00842000100041(10)10.3.0.2009(11)170309 Software Release PN 88-01319-00; Workstation PN: 88-01320-00 Version 10.3.1 UDI: (01) 00842000100041 (10) 10312135 (11) 171024 Application Server PN: 88-01426-00; Workstation PN 88-01430-00 Version 11.0.2 UDI: (01)00842000100416(10)11.0.2.1523(11)180809 Application Server Installation PN: 88-01570-00; Workstation Installation PN 88-01571-00 Version 11.0.3 UDI: (01)00842000100416(10)11.0.3.1601(11)181019 Workstation Installation PN: 88-01589-00 Version 11.1.1 UDI: (01)00842000100751(10)11.1.1(11)190724 Application Server 30-01650-00; Workstation PN:30-01654-00

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2403-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2403-2020. Report a data issue: hello@regdatalab.com.