Z-2404-2012 Class III Terminated

Recalled by Philips Medical Systems (Cleveland) Inc — Cleveland, OH

FDA device recall Z-2404-2012 was initiated by Philips Medical Systems (Cleveland) Inc on Aug 28, 2012 and is designated Class III. Reason for recall: Philips was notified that the system logout in software version 2.3.6 is now longer and may sometimes fail to logout, requiring a forced system restart or hard shutdown with the power button. The recall status is terminated (terminated Dec 24, 2013). Affected quantity: 93 units.

Recall details

Product Type
Devices
Report Date
Sep 26, 2012
Initiation Date
Aug 28, 2012
Termination Date
Dec 24, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
93 units

Product description

The Brilliance CT 16 System Product Usage: The Brilliance 16 is a Whole Body Computed Tomography X-ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles.

Reason for recall

Philips was notified that the system logout in software version 2.3.6 is now longer and may sometimes fail to logout, requiring a forced system restart or hard shutdown with the power button.

Distribution pattern

Worldwide Distribution - US Nationwide including the states of: AL, CA, GA, MA, MD, NC, OH, PA, TX, and WA; and to the following countries of: Algeria, Oman, Australia, Brazil, China, Saudi Arabia, United Arab Emirates, and India.

Code information

System Code #, 728246, Serial #, 6462, 6447, 6495, 6514, 6467, 6482, 6473, 6544, 6439, 6453, 6436, 6536, 6448, 6449, 6471, 6498, 6502, 6504, 6452, 6510, 6508, 6438, 6513, 6456, 6505, 6457, 6442, 6454, 6461, 6469, 6485, 6458, 6475, 6487, 6531, 6520, 6490, 6470, 6437, 6497, 6481, 6489, 6488, 6455, 6543, 6499, 6493, 6443, 6463, 6476, 6477, 6478, 6526, 6545, 6546, 6518, 6491, 6450, 6451, 6468, 6464, 6466, 6459, 6472, 6444, 6445, 6529, 6530, 6440, 6532, 6515, 6486, 6507, 6512, 6483, 6524, 6480, 6479, 6509, 6460, 6441, 6465, 6474, 6484, 6492, 6503, 6511, 6517, 6519, 6523, 6525, 6549, 6555

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2404-2012
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2404-2012. Report a data issue: hello@regdatalab.com.