Z-2416-2025 Class II Ongoing

Recalled by THOR Photomedicine Ltd — Amersham

Recall Details

Product Type
Devices
Report Date
September 3, 2025
Initiation Date
July 24, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
80 units (66 US, 14 OUS)

Product Description

Brand Name: THOR Product Name: NovoTHOR Model/Catalog Number: S2188 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: No

Reason for Recall

The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.

Distribution Pattern

Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom, Austria, Ireland, Italy, Lithuania, Netherlands, Poland, Serbia, Slovakia, Spain.

Code Information

Model/Catalog Number: S2188; UDI-DI: 05060494130503; Serial Numbers: 4004, 4015, 4017, 4019, 4021, 4022, 4024, 4026, 4028, 4029, 4035, 4036, 4037, 4042, 4043, 4044, 4046, 4047, 4049, 4054, 4055, 4059, 4066, 4071, 4072, 4073, 4074, 4076, 4077, 4078, 4079, 4082, 4083, 4084, 4085, 4086, 4087, 4088, 4090, 4091, 4094, 4095, 4097, 4099, 4101, 4105, 4106, 4111, 4112, 4114, 4116, 4117, 4121, 4123, 4124, 4125, 4126, 4130, 4140, 4141, 4143, 4144, 4151, 4153, 4157, 4158, 4009, 4020, 4103, 4107, 4109, 4110, 4113, 4120, 4122, 4127, 4133, 4134, 4138, 4139;