Z-2419-2025 Class II Ongoing
FDA device recall Z-2419-2025 was initiated by THOR Photomedicine Ltd on Jul 24, 2025 and is designated Class II. Reason for recall: The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift. The recall status is ongoing. Affected quantity: 2 units (1 US, 1 OUS).
Recall details
- Product Type
- Devices
- Report Date
- Sep 3, 2025
- Initiation Date
- Jul 24, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 units (1 US, 1 OUS)
Product description
Reason for recall
Distribution pattern
Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom, Austria, Ireland, Italy, Lithuania, Netherlands, Poland, Serbia, Slovakia, Spain.