Z-2419-2025 Class II Ongoing

Recalled by THOR Photomedicine Ltd — Amersham

FDA device recall Z-2419-2025 was initiated by THOR Photomedicine Ltd on Jul 24, 2025 and is designated Class II. Reason for recall: The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift. The recall status is ongoing. Affected quantity: 2 units (1 US, 1 OUS).

Recall details

Product Type
Devices
Report Date
Sep 3, 2025
Initiation Date
Jul 24, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units (1 US, 1 OUS)

Product description

Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2186 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: N/A

Reason for recall

The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.

Distribution pattern

Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom, Austria, Ireland, Italy, Lithuania, Netherlands, Poland, Serbia, Slovakia, Spain.

Code information

Model/Catalog Number: S2186; UDI-DI: Serial numbers: 4005, 4062;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2419-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2419-2025. Report a data issue: hello@regdatalab.com.