Z-2424-2018 Class II Terminated

Recalled by Ion Beam Applications S.A. — Louvain La Neuve

FDA device recall Z-2424-2018 was initiated by Ion Beam Applications S.A. on May 16, 2018 and is designated Class II. Reason for recall: Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position… The recall status is terminated (terminated Oct 16, 2018). Affected quantity: 3 units.

Recall details

Product Type
Devices
Report Date
Jul 18, 2018
Initiation Date
May 16, 2018
Termination Date
Oct 16, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units

Product description

Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Reason for recall

Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of FL and the countries of Sweden, The Netherlands

Code information

Serial Numbers: SAT.122 (EU) SAT.132 (EU) SAT.133 (US)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2424-2018
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2424-2018. Report a data issue: hello@regdatalab.com.