Z-2424-2018 Class II Terminated
FDA device recall Z-2424-2018 was initiated by Ion Beam Applications S.A. on May 16, 2018 and is designated Class II. Reason for recall: Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position… The recall status is terminated (terminated Oct 16, 2018). Affected quantity: 3 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 18, 2018
- Initiation Date
- May 16, 2018
- Termination Date
- Oct 16, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 units
Product description
Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
Reason for recall
Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.
Distribution pattern
Worldwide Distribution - US Nationwide in the states of FL and the countries of Sweden, The Netherlands
Code information
Serial Numbers: SAT.122 (EU) SAT.132 (EU) SAT.133 (US)