Z-2470-2018 Class II Terminated

Recalled by Ion Beam Applications S.A. — Louvain-la-neuve

FDA device recall Z-2470-2018 was initiated by Ion Beam Applications S.A. on Jun 27, 2018 and is designated Class II. Reason for recall: Gantry rotation or proton irradiation can be allowed even if apertures or accessories are not correctly positioned in snout 300x400 or snout XL. This can lead to risk of crushing for patient and/or… The recall status is terminated (terminated May 20, 2020). Affected quantity: 14.

Recall details

Product Type
Devices
Report Date
Jul 25, 2018
Initiation Date
Jun 27, 2018
Termination Date
May 20, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14

Product description

Proteus 235, Universal Nozzle with snout 300x400; PBS dedicated nozzle with snout XL Product Usage: Device is designed to produce and deliver proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.

Reason for recall

Gantry rotation or proton irradiation can be allowed even if apertures or accessories are not correctly positioned in snout 300x400 or snout XL. This can lead to risk of crushing for patient and/or user, and risk of mistreatment for the patient.

Distribution pattern

US Nationwide Distribution in the states of FL, VA, PA, OK, NJ, WA, and TX.

Code information

PAT.003, PAT.006, PAT.107, PAT.108, PAT.109, PAT.110, PAT.111, PAT.113, PAT.114, PAT.115, PAT.116, PAT.120, PAT.126, and PAT.132.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2470-2018
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2470-2018. Report a data issue: hello@regdatalab.com.