Z-2543-2023 Class II

Recalled by Landauer — Glenwood, IL

Recall Details

Product Type
Devices
Report Date
September 20, 2023
Initiation Date
July 12, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
535643 units

Product Description

nanoDot D2DNS, Model Numbers: a) 03056-000 (EMEA number VINLNAN002); b) 03056-1MO; c) 03056-3MO; d) 03056-KIT; radiation monitoring dosimeter used with the microSTAR readers

Reason for Recall

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Distribution Pattern

Worldwide

Code Information

UDI/DI 0860003399903, all batch numbers