Z-2555-2018 Class II Terminated

Recalled by Leica Microsystems, Inc. — Buffalo Grove, IL

FDA device recall Z-2555-2018 was initiated by Leica Microsystems, Inc. on Jun 4, 2018 and is designated Class II. Reason for recall: Possible failure of screws which hold the gas spring fixation bracket in place. Due to this failure, the gas spring, which is fixed to the upper part of the parallelogram, cannot hold up the weight… The recall status is terminated (terminated Mar 25, 2021). Affected quantity: 8,564 Total Units.

Recall details

Product Type
Devices
Report Date
Aug 1, 2018
Initiation Date
Jun 4, 2018
Termination Date
Mar 25, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8,564 Total Units

Product description

Leica Microsystems products labeled as the following: Leica M320 Microscope Systems with Article Numbers 10448419, 10448420, 10448485, 10448486 Product Usage: Intended use of this device is an optical instrument for magnifying and illuminating specimens. It can be applied for observation and documentation and for human and veterinary medical treatment. Not for use in Eye Operations.

Reason for recall

Possible failure of screws which hold the gas spring fixation bracket in place. Due to this failure, the gas spring, which is fixed to the upper part of the parallelogram, cannot hold up the weight of the optics carrier.

Distribution pattern

Worldwide Distribution US Nationwide in the states of : CA, CO, CT, FL, HI, ID, IL, KS, KY, LA, MD, MI, MN, MO, NC, NJ, NY, OH, PA, SD, TN, TX. International Distribution: Canada, AUSTRALIA, BELGIUM, Bolivia, BRAZIL, BULGARIA, CHILE, CHINA, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, HONGKONG, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KUWAIT, MALAYSIA, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, TAIWAN,THAILAND, TURKEY, UK

Code information

A. Article Number 10448419 (Serial Numbers 160714001, 10814001, 200814001, 200814002, 90914001, 021014001, 21214001, 31214001, 81214001, 190115001, 190115002); B Article Number 10448420 (Serial Numbers 180514001, 180514002, 190514001, 190514002, 190514003, 190514004, 200514001, 200514002, 200514003, 200514004, 210514001, 210514002, 210514003, 210514004, 220514001, 220514002, 220514003, 260514001, 260514002, 270514001, 270514002, 270514003, 270514004, 270514005, 280514001, 280514002, 280514003, 290514001, 300514001, 300514002, 300514003, 20614001, 40614001, 50614001, 50614002, 50614003, 60614001, 90614001, 90614002, 100614001, 100614002, 100614003, 120614001, 120614002, 120614003, 130614001, 130614002, 130614003, 130614004, 90514002, 160614001, 160614002, 160614003, 160614004, 170614001, 190614001, 200614001, 200614002, 200614003, 230614001, 230614002, 230614003, 240614001, 250614001, 250614002, 250614003, 250614004, 250614005, 270614001, 300614001, 300614002, 300614003, 10714001, 10714002, 10714003, 10714004, 20714001, 20714002, 20714003, 30714001, 30714002, 40714001, 40714002, 40714003, 70714001, 70714002, 70714003, 80714001, 80714002, 80714003, 90714001, 100714001, 110714001, 140714001, 160714001, 160714002, 170714001, 170714002, 180714001, 180714002, 180714003, 210714001, 210714002, 210714003, 220714001, 220714002, 220714003, 220714004, 230714001, 250714001, 250714002, 250714003, 250714004, 250714005, 250714006, 290714001, 290714002, 290714003, 290714004, 300714001, 300714002, 300714003, 300714004, 310714001, 310714002, 310714003, 310714004, 10814001, 10814002, 040814001, 040814002, 040814003, 040814004, 050814001, 050814002, 050814003, 050814004, 060814001, 060814002, 070814001, 070814002, 070814003, 070814004, 080814001, 080814002, 120814001, 120814002, 130814001, 130814002, 130814003, 130814004, 140814001, 140814002, 140814003, 140814004, 150814001, 150814002, 150814003, 160814001, 160814002, 160814003, 180814001, 180814002, 180814003, 190814001, 190814002, 190814003, 190814004, 190814005, 190814006, 200814001, 200814002, 210814001, 210814002, 210814003, 210814004, 220814001, 220814002, 220814003, 250814001, 250814002, 270814001, 270814002, 270814003, 270814004, 270814005, 270814006, 280814001, 280814002, 280814003, 280814004, 290814001, 290814002, 10914001, 10914002, 10914003, 10914004, 20914001, 20914002, 30914001, 30914002, 30914003, 80714002, 70814003, 40914001, 40914002, 50914001, 50914002, 80914001, 80914002, 80914003, 80914004, 80914005, 90914001, 90914002, 90914003, 100914001, 100914002, 100914003, 100914004, 110914001, 110914002, 110914003, 120914001, 120914002, 120914003, 120914004, 120914005, 130914001, 130914002, 130914003, 150914001, 160914001, 160914002, 160914003, 160914004, 160914005, 170914001, 170914002, 170914003, 170914004, 170914005, 180914001, 180914002, 180914003, 180914004, 180914005, 190914001, 190914002, 190914003, 200914001, 200914002, 200914003, 200914004, 200914005, 220914001, 220914002, 220914003, 220914004, 230914001, 230914002, 90614001, 90614002, 240914001, 240914002, 240914003, 250914001, 250914002, 100614002, 100614003, 260914001, 260914002, 260914003, 260914004, 290914001, 290914002, 290914003, 300914001, 300914002, 300914003, 021014001, 021014002, 031014001, 031014002, 031014003, 031014004, 41014001, 41014002, 41014003, 41014004, 41014005, 41014006, 41014007, 61014001, 61014002, 71014001, 71014002, 71014003, 71014004, 81014001, 81014002, 81014003, 81014004, 81014005, 91014001, 91014002, 101014001, 101014002, 101014003, 101014004, 121014001, 121014002, 121014003, 121014004, 121014005, 121014006, 131014001, 141014001, 141014002, 141014003, 141014004, 151014001, 151014002, 151014003, 151014004, 161014001, 161014002, 161014003, 171014001, 171014002, 171014003, 171014004, 181014001, 181014002, 181014003, 181014004, 191014001, 191014002, 191014003, 191014004, 201014001, 201014002, 201014003, 211014001, 211014002, 211014003, 211014004, 231014001, 231014002, 231014003, 241014001, 241014002, 241014003, 241014004, 241014005, 281014001, 281014002, 281014003, 281014004, 281014005, 291014001, 291014002, 291014003, 291014004, 291014005, 291014006, 301014001, 301014002, 311014001, 311014002, 311014003, 311014004, 11114001, 11114002, 11114003, 11114004, 11114005, 31114001, 31114002, 31114003, 31114004, 31114005, 41114001, 41114002, 41114003, 51114001, 51114002, 51114003, 51114004, 81114001, 81114002, 81114003, 81114004, 101114001, 101114002, 101114003, 111114001, 111114002, 121114001, 121114002, 121114003, 121114004, 131114001, 131114002, 141114001, 141114002, 141114003, 161114001, 161114002, 161114003, 161114004, 161114005, 161114006, 171114001, 171114002, 171114003, 171114004, 181114001, 181114002, 181114003, 191114001, 191114002, 191114003, 201114001, 201114002, 201114003, 201114004, 201114005, 211114001, 211114002, 211114003, 211114004, 241114001, 241114002, 241114003, 241114004, 251114001, 251114002, 251114003, 251114004, 261114001, 261114002, 261114003, 261114004, 271114001, 271114002, 281114001, 281114002, 11214001, 11214002, 21214001, 31214001, 31214002, 31214003, 41214001, 41214002, 41214003, 51214001, 51214002, 71214001, 71214002, 81214001, 81214002, 81214003, 91214001, 91214002, 91214003, 91214004, 101214001, 101214002, 101214003, 111214001, 111214002, 111214003, 121214001, 121214002, 121214003, 121214004, 151214001, 151214002, 151214003, 151214004, 151214005, 151214006, 151214007, 151214008, 161214001, 161214002, 171214001, 171214002, 171214003, 181214001, 181214002, 181214003, 181214004, 191214001, 191214002, 191214003, 191214004, 191214005, 191214006, 211214001, 221214001, 221214002, 241214001, 241214002, 241214003, 261214001, 261214002, 261214003, 261214004, 271214001, 271214002, 271214003, 281214001, 281214002, 281214003, 20115001, 20115002, 20115003, 20115004, 50115001, 50115002, 60115001, 60115002, 60115003, 60115004, 60115005, 60115006, 70115001, 70115002, 80115001, 80115002, 90115001, 90115002, 90115003, 90115004, 110115001, 110115002, 120115002, 120115003, 120115004, 130115001, 130115002, 130115003, 130115004, 140115001, 140115002, 140115003, 140115004, 150115001, 150115002, 160115001, 160115002, 190115001, 190115002, 200115001, 200115002, 200115003, 200115004, 210115002, 210115004, 130215004, 110515005, 120115004, 130115003, 50915001); C. Article Number 10448485 (Serial Numbers 180514001, 180514002, 230514001, 230514002, 230514003, 230514004, 260514001, 260514002, 280514001, 290514001, 20614001, 20614002, 20614003, 30614001, 30614002, 40614001, 40614002, 60614001, 60614002, 90614001, 90614002, 110614001, 110614002, 170614001, 170614002, 170614003, 170614004, 170614005, 180614001, 190614001, 190614002, 190614003, 190614004, 190614005, 200614001, 230614001, 300614001, 20714001, 30714001, 30714002, 40714001, 80714001, 90714001, 90714002, 100714001, 100714002, 110714001, 110714002, 140714001, 140714001, 170714001, 180714001, 210714001, 080814001, 080814002, 120814001, 120814002, 150814001, 220814001, 260814001, 260814002, 260814003, 260814004, 290814001, 290814002, 20914001, 20914002, 50914001, 50914002, 110914001, 130914001, 150914001, 150914002, 230914001, 240914001, 250914001, 250914002, 290914001, 300914001, 011014001, 011014002, 011014003, 011014004, 021014001, 41014001, 61014001, 91014001, 91014002, 101014001, 121014001, 151014001, 211014001, 231014001, 241014001, 241014002, 301014001, 301014002, 301014003, 311014001, 11114001, 11114002, 11114003, 41114001, 81114001, 101114001, 111114001, 111114002, 191114001, 211114001, 271114001, 271114002, 281114001, 11214001, 21214001, 21214002, 51214001, 71214001, 71214002, 71214003, 101214001, 111214001, 161214001, 161214002, 171214001, 191214001, 191214002, 221214001, 221214002, 231214001, 231214002, 231214003, 241214001, 251214001, 251214002, 251214003, 70115001, 80115001, 80115002, 150115001, 150115002, 160115001, 160115002); D. Article Number 10448486 (Serial Numbers 110614001, 110614002, 120614001, 310114001, 060814001, 060814002, 160814001, 61014001, 121014001, 281014001, 131114001, 131114002, 181114001, 231214001)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2555-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Leica Microsystems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Leica Microsystems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Leica Microsystems, Inc. →

Source: openFDA record for Z-2555-2018. Report a data issue: hello@regdatalab.com.