Z-2560-2019 Class II Ongoing

Recalled by Elekta Inc — Maryland Heights, MO

FDA device recall Z-2560-2019 was initiated by Elekta Inc on Sep 4, 2019 and is designated Class II. Reason for recall: It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation. The recall status is ongoing. Affected quantity: 30 units.

Recall details

Product Type
Devices
Report Date
Oct 2, 2019
Initiation Date
Sep 4, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
30 units

Product description

Monaco RTP System, 5.40 Unity, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review

Reason for recall

It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of WI, OH, NY, OH, IN, IA, TX, and countries of Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, UK

Code information

UDI (01)00858164002190(10) 5.40.00

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2560-2019
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2560-2019. Report a data issue: hello@regdatalab.com.