Z-2562-2023 Class II
Recall Details
- Product Type
- Devices
- Report Date
- September 20, 2023
- Initiation Date
- July 12, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 211815 units
Product Description
nanoDot D2DNS, Model Numbers: a) 18100-000; b) 18100-1MO; c) 18100-2MO; d) 18100-KIT; radiation monitoring dosimeter used with the microSTAR readers
Reason for Recall
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Distribution Pattern
Worldwide
Code Information
UDI/DI 0860003399903, all batch numbers