Z-2580-2016 Class III Terminated

Recalled by ORTHO-CLINICAL DIAGNOSTICS — Bridgend, N/A

FDA device recall Z-2580-2016 was initiated by ORTHO-CLINICAL DIAGNOSTICS on Jul 28, 2016 and is designated Class III. Reason for recall: A field correction was issued by the firm for VITROS TSH reagent lot 5040 due to a higher than expected frequency of customer calibration curves falling outside of the calibration quality parameters… The recall status is terminated (terminated Jun 11, 2018). Affected quantity: Domestic: 2883, Foreign: 1261.

Recall details

Product Type
Devices
Report Date
Aug 24, 2016
Initiation Date
Jul 28, 2016
Termination Date
Jun 11, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Domestic: 2883, Foreign: 1261

Product description

VITROS Immunodiagnostic Products TSH Reagent Pack, REF/Catalog Number/Product Code 191 2997, Unique Device Identifier No. 10758750000227, IVD --- Ortho-Clinical Diagnostics Pencoed Bridgend CF35 5PZ UK For the in vitro quantitative measurement of thyroid stimulating hormone (TSH) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System to aid in the differential diagnosis of thyroid disease.

Reason for recall

A field correction was issued by the firm for VITROS TSH reagent lot 5040 due to a higher than expected frequency of customer calibration curves falling outside of the calibration quality parameters and therefore unusable to process samples. Customers using VITROS TSH Lot 5040 will potentially experience an unexpected increase in the unsuccessful calibrations potentially causing a delay in testing and reporting results for VITROS TSH.

Distribution pattern

Worldwide Distribution - US to VA, WV, GA, TN, KY, MN, SD, AR, OK and Internationally to Canada

Code information

Lot 5040 (Expiry date 19 Dec 2016)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)