Z-2581-2014 Class II Terminated

Recalled by Elekta, Inc. — Atlanta, GA

FDA device recall Z-2581-2014 was initiated by Elekta, Inc. on Aug 18, 2014 and is designated Class II. Reason for recall: Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported… The recall status is terminated (terminated Sep 15, 2021). Affected quantity: 928 units.

Recall details

Product Type
Devices
Report Date
Sep 10, 2014
Initiation Date
Aug 18, 2014
Termination Date
Sep 15, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
928 units

Product description

Monaco The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and displays, on-screen and in hard-copy, two or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups.

Reason for recall

Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan.

Distribution pattern

Worldwide distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, FL, IL, IN, MD, MA, MI, MN, MO, MT, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, Puerto Rico; and countries of: Algeria, Angola, Antigua & Barbuda, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sir Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Ukraine.

Code information

Versions 3.20.00 and higher

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2581-2014
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2581-2014. Report a data issue: hello@regdatalab.com.