Z-2624-2026 Class I Ongoing

Recalled by Medtronic Heart Valves Division — Santa Ana, CA

FDA device recall Z-2624-2026 was initiated by Medtronic Heart Valves Division on May 22, 2026 and is designated Class I. Reason for recall: Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical… The recall status is ongoing. Affected quantity: 1881 units.

Recall details

Product Type
Devices
Report Date
Jul 8, 2026
Initiation Date
May 22, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1881 units

Product description

Medtronic Harmony Delivery Catheter System

Reason for recall

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Canada, Chile, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Mexico, Netherlands, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Code information

GTIN 00763000341367, Lot Numbers: 0012508285, 0012517045, 0012526102, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663878, 0012663879, 0012690218, 0012699635, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824770, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965210, 0012965211, 0012976400, 0012986693, 0013063766, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624, 0013236643, 0013272796, 0013272798. GTIN 00763000544027, Lot Numbers: 0012508285, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663879, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965211, 0012976400, 0012986693, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624. GTIN 00763000520151, Lot Numbers: 0012575897, 0012642965, 0012728368, 0013272799. GTIN 00763000582951, Lot Numbers: 0012590535, 0012606411, 0012606413, 0012642964, 0012659418, 0012678282, 0012690219, 0012699637, 0012804151, 0012804580, 0012874527, 0012888682, 0012896885, 0012965221, 0012986699, 0013063764, 0013063765, 0013105037, 0013227504. GTIN 00763000918712, Lot Numbers: 0012635392, 0012676200, 0012690207, 0012874528, 0012874529, 0013015539, 0013105038.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2624-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2624-2026. Report a data issue: hello@regdatalab.com.